![Rohde & Schwarz R&S NRP Operating Manual Download Page 191](http://html.mh-extra.com/html/rohde-and-schwarz/rands-nrp/rands-nrp_operating-manual_1477868191.webp)
Remote Control - Commands
R&S NRP
1144.1400.12 6.16
E-3
Calculation Functions
The CALCulate blocks receive measurement data from the sensors via input channels which are called
primary and secondary channel. The following functions are available for calculating the primary
channel (PC) and the secondary channel (SC).
One-parameter functions (
<source_list>
=
(@
n
)
;
n
=
1,2,3,4
)
-
PC
Yields the measured average power of the sensor assigned to
the PC.
:RELative
PC / reference value
The value measured in the PC is divided by the reference value
determined with the aid of one of the following commands:
CALCulate[1..8]:RELative[:MAGNitude]
CALCulate[1..8]:RELative:AUTO
.
Two-parameter functions (
<source_list>
=
(@
n
),(@
m
)
;
n
=
1,2,3,4;
m
=
1,2,3,4
)
:DIFFerence
PC - SC
Yields the difference between the PC and SC. This means that
two sensors must be specified in
<source_list>.
This also
applies to the following functions.
:DIFFerence:RELative
(PC - SC) / reference value
The difference between the PC and SC is divided by the
reference value determined with the aid of one of the following
commands:
CALCulate[1..8]:RELative[:MAGNitude]
CALCulate[1..8]:RELative:AUTO
.
:SUM
PC+ SC
Yields the sum of the PC and SC. The
<source_list>
must
therefore contain two sensors. This also applies to the following
functions.
:SUM:RELative
(PC + SC) / reference value
The sum of the PC and SC is divided by the reference value
determined with the aid of one of the following commands:
CALCulate[1..8]:RELative[:MAGNitude]
or
CALCulate[1..8]:RELative:AUTO
.
:RATio
PC / SC
Yields the ratio of PC to SC.
:RATio:RELative
(PC / SC) / reference value
The ratio of the PC and SC is divided by the reference value
determined with the aid of one the following commands:
CALCulate[1..8]:RELative[:MAGNitude]
CALCulate[1..8]:RELative:AUTO
.