
SCPI Error Messages
R&S NRP
1144.1400.12
9.6
E-2
Error code
Error text returned upon queue query
Explanations
28
Sensor not idle
A command could not be executed because the respective sensor was not in the IDLE state at that time.
29
Measurement aborted while waiting for data
A measurement was aborted while the device was waiting for measurement results.
30
No extremes available in this mode
Extreme values (minimum, maximum und peak-to-peak values) are not calculated in the set measurement mode.
39
Sensor removed while waiting for result
A sensor was disconnected while the device was waiting for measurement results.
42
Failed to initialize sensor
A sensor could not be initialized.
43
Error in receiving calibration data from sensor
An error occurred in receiving calibration data from the sensor (command
CALibration[1..4]:DATA?
).
44
Error in sending calibration data to sensor
An error occurred in sending calibration data to the sensor
(command
CALibration[1..4]:DATA <block_data>
).
45
Command not supported by sensor
Attempt was made to send a command to a sensor that does not support this command. This may depend on the
current sensor status. The sensors can inform the R&S NRP at any point which commands they support or do not
support. This mainly concerns commands of the Sense and Trigger systems.
46
Sensor failure: no command receipt
A sensor did not acknowledge any command.
47
Error in receiving battery data
An error occurred in receiving battery data (command
SYSTem:BATTery:STATus?
).
48
Error in sending battery data
An error occurred in sending battery data (command
SYSTem:BATTery:MODE <block_data>
).
50
Fatal sensor error
A sensor signalled a serious error. If the problem persists, please contact the R&S servicing center.
51
Overload
A sensor signalled that the signal applied has exceeded the permissible maximum power. CAUTION! Sensor
overload can destroy the electronic measurement system.
52
Overrange
This error is signalled if a sensor detects that one of its measurement channels is overranged. The result is then
probably incorrect. This error can occur if the automatic range function is deactivated (
SENSe:RANGe:AUTO
OFF
) and the manually selected measurement channel is not suitable or the cross-over level
(
SENSe:RANGe:AUTO:CLEVel
) was incorrectly selected.
53
Truncated measurement
In the BurstAv mode, the samples for the power are stored in an internal buffer until the end of the power pulse is
detected. If the buffer overflows the measurement is terminated and error 53 "Truncated measurement" is
generated.