MEAS:ARRay? (1000)
1000 measured values are recorded in the ContAv mode. After termination of the last measurement,
the results are stored in the output queue. There is no triggering as in the ContAv mode. The
measurement result is a list of 1000 measured values separated by a comma.
The keywords for processing two sensors can be added to the
MEAS:TSL
,
MEAS:BURS
and
MEAS:ARRay
commands, but the
:SWR
,
:RLOS
and
:REFL
functions are not available.
Example:
MEAS:TSL:RAT? 577 us, 8, 18 us, 18 us, DEF, 3, (@1), (@3)
The above timeslot example is extended to two sensors. The measurement result is a list of 8
measured values separated by a comma for the 8 timeslots; these values were obtained by division
from the measured values of sensors 1 and 3.