R&S NRP
Annex: Remote Control - Commands
1144.1400.12
Annex - I.21
E-2
Optimization of measurement speed
The duration of a message can be reduced by making the following setting:
SYST:SPEed FAST
(deactivates screen content !!!)
The following settings also affect the measurement speed to a greater or lesser extent. Which of these
settings are possible, however, depends on what is to be measured. For example, if measurement
accuracy is important, filtering may also have to be activated, which can reduce the measurement
speed. The length of the integration time also directly affects the measurement time and should
therefore be kept as short as possible if a short measurement time is desired. Consequently, the
following list can only serve as a guideline for minimizing the measurement time.
SYST:SPEed FAST
(deactivates screen content)
SENS:AVER:STAT OFF
(deactivates filtering of measured values)
SENS:FUNC "POW:AVG"
(ContAv mode)
SENS:BUFF:STAT OFF
(no buffered measurements)
SENS:APER MIN
(integration time as short as possible)
SENS:RANG:AUTO OFF
(automatic range selection OFF)
SENS:CORR:OFFS:STAT OFF
(no global offset correction)
SENS:CORR:DCYC:STAT OFF
(no duty cycle correction)
SENS:CORR:FDOT:STAT OFF
(no frequency-dependent offset correction)
SENS:SGAM:CORR:STAT OFF
(no measured-value correction for reflection coefficient of source)
TRIG:DEL 0
(no delay on triggering)
TRIG:DEL:AUTO OFF
(no delay on triggering)
CALC1:MATH "(SENS1)"
(only uses the first CALCulate block)
CALC2:MATH "(SENS2)"
CALC3:MATH "(SENS2)"
CALC4:MATH "(SENS2)"
CALC5:MATH "(SENS2)"
CALC6:MATH "(SENS2)"
CALC7:MATH "(SENS2)"
CALC8:MATH "(SENS2)"