R&S NRP
Remote Control - Commands
1144.1400.12 6.41
E-3
MEMory:FREE:STATe?
Returns the number of bytes available in memory for storing device states as well as the memory
already used for device states.
MEMory:FREE:TABLe?
Returns the number of bytes available in memory for storing device states as well as the memory
already used for device states.
MEMory:NSTates?
Returns the number of storable device states. Since 20 device states can be stored, 20 is always
returned.
MEMory:STATe:CATalog?
Returns a list of names of stored device states in the format
<string>{,<string>}
MEMory:STATe:DEFine|:MAP <string>, 1..19
MEMory:STATe:DEFine|:MAP? <string>
The value 0 to 19 are assigned to the memory locations of the device states, where the memory
location with the number 0 is the factory-set state and cannot be changed. Names
(<string>)
can be
assigned to these numbers with
MEM:STAT:DEF
(except for the factory-set state). The command
MEM:CLE[:NAME]
expects parameters that are actually strings rather than the numbers of the memory
locations. The default names of these memory locations are "Setup 0" to "Setup 19". The name may
contain the characters A to Z, a to z, 0 to 9 and "_".
The query of this command returns the number of the memory location with the name of
<string>
.
MEMory:TABLe:FREQuency[?] <NRf>{,<NRf>}
This command enters frequencies in the first column of the table selected with
MEM:TABL:SEL
.
Existing data for frequencies will be overwritten. The number of frequencies should match the number
of offset values. If the number of frequencies differs from the number of offset values, the shorter
column length will be used.
SENS[1..4]:FREQ
can be used to determine a frequency for which a correction factor is then defined by
means of the offset table. If the exact frequency selected is not in the table, the values in the table are
interpolated. If the selected frequency is outside the specified frequency range, the frst or last offset
value of the table is selected. Interpolation is linear in the units
HZ
and
DB
.
Unit:
HZ
*RST values
:
Tables are not changed in case of a
*RST
.