
Switching the meter on/off
R&S NRP
1144.1400.12 1.10
E-2
If an error is detected when the non-volatile memory contain-
ing instrument settings is checked, an error message is dis-
played when booting is completed. The non-volatile memory
is then re-initialized completely and the R&S NRP goes into
the preset state. All instrument settings that have been
saved are lost as a result of this procedure.
Resetting and setting brightness and contrast
If you press the
(Æ)
key immediately after switching on the
R&S NRP, after about 3 seconds the R&S NRP goes into a
state in which you can adjust brightness and contrast.
Initially, contrast and brightness are automatically set to
their default values. You can then make the settings you
want using the diagram on the screen.
ON state
Whenever the meter is switched off, its settings are saved. The next time it is switched on, these set-
tings are automatically restored.
If the last setup before switch-off is to be restored, the R&S NRP must be switched off
with the standby key
( Ø )
. If the meter is turned off by disconnecting the AC supply, it is
possible that the settings being used immediately beforehand are not saved.