R&S NRP
Configuring measurements (Measurement menu)
1144.1400.12 4.35
E-2
Expanded configuration for sensors in the Scope mode
The measurement results of sensors operated in the Scope mode can be further processed and dis-
played in two ways:
a) by a power/time diagram (measurement window in the Graph display mode
Selecting the result
display mode, p. 4.29).
b) in the form of gated measurement results (measurement window in the Dig or D&A display)
Whereas the Graph mode allows only an absolute display of the measured values, gated measurement
results can also be further processed like other measurement results: as absolute values relative to a
fixed reference value or another measurement channel. For this purpose, the
Func&Unit
dialog box
provides expanded configuration possibilities, including the number of the gate and the type of meas-
urand (Avg, Peak, Peak/Avg).
Note the following:
•
Restrictions regarding Peak and Peak/Avg are described in the background information to
Meas, p. 4.47
•
Only functions whose result is a power or a dimensionless ratio can be created
•
The reference of a gate measured value to the value of another gate in the same channel is not
possible
Func & Unit- Dialog
Primary or secondary
channel in the Scope
mode.
Fig. 4-31
Func & Unit dialog box
Gate# –
( )
ä
↔
ä
Determines which gate is to be evaluated for the function.
Selects one of the available measurands:
Avg
Average power in the gate
Peak
Peak power in the gate
Eval. –
( )
ä
↔
ä
Peak/Avg
Ratio of peak power to average power in the gate