Configuring measurements (Measurement menu)
R&S NRP
1144.1400.12 4.38
E-2
Setting the auxiliary value
In the Digital display mode, the result and another value that character-
izes the result can be displayed.
Fig. 4-32 Measurement window,
Auxiliary Value
Selecting the auxiliary value:
Max
The greatest value that has been measured since
the extreme value function was last reset
(
Resetting the extreme value display, page
4.38).
Min
The smallest value that has been measured
since the extreme value function was last reset
(
Resetting the extreme value display, page
4.38).
Auxiliary Value
–
( )
ä
↔
ä
Max-Min
The interval that includes all values measured
since the extreme value function was last reset
(
Resetting the extreme value display, page
4.38).
Remote Control
CALCulate[1..8][:MINimum | :MAXimum | :PTPeak]:DATA?
DISPlay[:WINDow][1..4]:AVALue[?] MINimum|MAXimum|PTPeak|OFF
Resetting the extreme value display
For each window, the R&S NRP continuously saves the maximum and minimum value measured since
the start of the measurement or since these values were last reset. These values should, therefore, be
reset before the measurement is started.
Measurement menu
:
Max Hold Reset
–
( )
ä
Clears the extreme values and restarts recording of extreme values.
Remote Control
CALCulate[1..8]:EXTRemes:RESet
Func & Unit : Auxiliary Value
Measurement
Max Hold Reset