R&S NRP
Remote Control - Commands
1144.1400.12 6.59
E-3
[SENSe[1..4]][:POWer]:TGATe[1..4]:OFFSet:TIME[?] <float_value>
This command is used for configuring the Timegate mode. It specifies the start of one of the four gates
(sweeps) relative to the delayed trigger point. If this value is negative, the gate starts before the trigger
point. This is possible because the sensor stores samples of the measured powers and can thus
transfer the start of a measurement to the past. In manual operation, the start and duration of the gates
can also be defined graphically if the sensor is operated in the Scope mode.
Unit:
S
Value range:
depending on sensor
*RST
value:
depending on sensor
Trigger level
Power
Time
Trigger event
Delayed
trigger point
Trigger
delay
Gate 1
Gate 4
Gate 3
Gate 2
SENS:TRAC1:OFFS:TIME SENS:TRAC1:TIME
SENS:TRAC2:OFFS:TIME
SENS:TRAC2:TIME
Figure 6-8: Timing parameters for the Timegate mode.
[SENSe[1..4]][:POWer]:TGATe:SELect[?] 1 | 2 | 3 | 4
This command selects one of the four gates in which the average power is measured if the Timegate
mode is selected.
*RST
value:
1
[SENSe[1..4]][:POWer]:TGATe[1..4]:TIME[?] <float_value>
This command is used for configuring the Timegate mode. It specifies the duration of one of the four
gates (sweeps) relative to the delayed trigger point. In manual operation, the start and duration of the
gates can also be defined graphically if the sensor is operated in the Scope mode.
Unit:
S
Value range:
depending on sensor
*RST
value:
depending on sensor