![Rohde & Schwarz R&S NRP Operating Manual Download Page 226](http://html.mh-extra.com/html/rohde-and-schwarz/rands-nrp/rands-nrp_operating-manual_1477868226.webp)
R&S NRP
Remote Control - Commands
1144.1400.12 6.51
E-3
Command
Parameter
Unit
Remark
:BURSt
:DTOLerance[?]
<float_value>
S
:SAMPLing[?]
FREQ1|FREQ2
-
:SGAMma
[:MAGNitude][?]
<float_value>
-
:PHASe[?]
<float_value>
-
:CORRection:STATe[?]
ON|OFF
-
:TRACe
:AVERage
[:STATe][?]
ON|OFF
-
:COUNt[?]
<int_value>
-
:AUTO[?]
ON|OFF|ONCE
-
:MTIMe[?]
<float_value>
S
:RESolution[?]
1|2|3|4
-
:POINt[?]
<int_value>
-
:NSRatio[?]
<float_value>
DB|PCT
:TYPE[?]
RESolution|NSR
-
:TCONtrol[?]
MOVing|REPeat
-
:OFFSet:TIME[?]
<float_value>
S
:POINts[?]
<int_value>
-
:REALtime[?]
ON|OFF
-
:TIME[?]
<float_value>
S
:TIMing:EXCLude
:STARt[?]
<float_value>
S
:STOP[?]
<float_value>
S
[SENSe[1..4]]:AVERage[:STATe][?] ON | OFF
This command switches the filter function of a sensor on or off. When the filter is switched on, the
number of measured values set with
[
SENSe[1..4]]:AVERage:COUNt
is averaged. This reduces
the effect of noise so that more reliable results are obtained.
*RST value
:
depending on sensor
[SENSe[1..4]]:AVERage:COUNt[?] <int_value>
This command sets the filter length i.e. the number of readings to be averaged for one measured
value.. The higher the count the lower the noise and the longer it takes to obtain a measured value.
Unit:
-
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:AVERage:COUNt[?]:AUTO[?] ON | OFF | ONCE
This command can be used to automatically determine a value for the filter legth
[SENSe[1..4]]:AVERage:COUNt
. If the command is called with the parameter
ONCE,
the filter
length is automatically detected during the next measurement and set to that value. Thereafter
automatic filter length detection is switched off. The setting can then be queried via
SENS[1..4]:AVER:COUN?
. If the automatic switchover is activated with the ON parameter, the sensor
always defines a suitable filter length which can also be queried via
SENS[1..4]:AVER:COUN?
.
*RST value
:
depending on sensor