![Rohde & Schwarz R&S NRP Operating Manual Download Page 166](http://html.mh-extra.com/html/rohde-and-schwarz/rands-nrp/rands-nrp_operating-manual_1477868166.webp)
Remote Control - Fundamentals R&S
NRP
1144.1400.12 5.16
E-3
Mode
SCPI command
Value
Remark
TRIGger[1..4]:SOURce
EXTernal | INTernal
User-selectable.
TRIGger[1..4]:SLOPe
POSitive | NEGative
User-selectable.
TRIGger[1..4]:COUNt 1
Cannot be modified in manual
mode.
TRIGger[1..4]:DELay
Depends on sensor
User-selectable.
TRIGger[1..4]:HOLDoff
Depends on sensor
User-selectable.
TRIGger[1..4]:LEVel
Depends on sensor
User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:STARt -
Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:STOP -
Unchanged, is irrelevant.
Timeslot
INITiate[1..4]:CONTinuous ON
Cannot be modified in manual
mode.
TRIGger[1..4]:SOURce
EXTernal | INTernal
User-selectable.
TRIGger[1..4]:SLOPe
POSitive | NEGative
User-selectable.
TRIGger[1..4]:COUNt 1
Cannot be modified in manual
mode.
TRIGger[1..4]:DELay 0
S
Cannot be modified in manual
mode.
TRIGger[1..4]:HOLDoff
Depends on sensor
User-selectable.
TRIGger[1..4]:LEVel
Depends on sensor
User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:STARt 0
S
Cannot be modified in manual
mode.
[SENSe[1..4]]:TIMing:EXCLude:STOP 0
S
Cannot be modified in manual
mode.
Scope
INITiate[1..4]:CONTinuous ON
Cannot be modified in manual
mode.