R&S NRP
Configuring measurements (Measurement menu)
1144.1400.12 4.39
E-2
Additional functions
Relative measurements
The
Rel
function is a simple way of obtaining the ratio of a measured value to a constant value, e.g. a
measured power.
Measurement menu
:
Rel –
( )
Off
On Reset
ä
↔
ä
Off:
Measurement using the measurement function that has been set.
Rel –
( )
Off
On
Reset
ä
↔
ä
On:
Relative display mode.
Remote Control
CALCulate[1..8]:RELative:STATe[?] ON | OFF
Rel –
( )
Off On
Reset
ä
↔
ä
If the right-hand side of the rocker switch is pressed when
On
is se-
lected, the current measured value is transferred to the active window’s
reference memory. The display then returns to
On
.
Fernsteuerung
CALCulate[1..8]:RELative[:MAGNitude]:AUTO[?] ONCE
Setting the reference value manually
The reference value can also be directly entered or edited using the
Ref. Val…
dialog box.
Measurement menu
:
Ref. Val…
–
( )
ä
Fig. 4-33
Ref. Value dialog box
Ref. Value
–
( )
ä
ä
For directly entering a new reference value or for editing an old value.
Remote Control
CALCulate[1..8]:RELative[:MAGNitude][?] <float_value>
Result
Ref
–
( )
ä
ä
The current result is transferred to the reference value memory of the
active window and displayed. This function can also be started from
the Measurement menu by using the
Reset
softkey.
Remote Control
CALCulate[1..8]:RELative[:MAGNitude]:AUTO[?]
ONCE
Measurement
Rel
Measurement
Ref. Val