
R&S NRP
Remote Control - Commands
1144.1400.12 6.137
E-3
"WCDMA/3GPP TDD UL"
Sensor parameter
Remote control command in short form
Value
Sensor Mode
SENS:FUNC "POW:XTIM"
Integration Time
SENS:APER
10 ms
Duty Cycle
SENS:DCYC
6.420 %
Duty Cycle Correction
SENS:DCYC:STAT ON
Smoothing
SENS:SMO OFF
Dropout Tolerance
SENS:BURST:DTOL
666.667 µs
Exclude Time Start
SENS:TIM:EXCL:STAR
15 µs
Exclude Time End
SENS:TIM:EXCL:STOP
40 µs
Number Of Timeslots
SENS:TSL:COUN
15
Timeslot Width
SENS:TSL:WIDT
666.667 µs
Timegate Offset
SENS:TGAT:OFFS:TIME
1: 15 µs
2: 5.348 ms
Timegate Time
SENS:TGAT:TIME
1: 611.667 µs
2: 612 µs
Scope Offset Time
SENS:TRAC:OFFS:TIME
-200 µs
Scope Capture Time
SENS:TRAC:TIME
10.250 ms
Scope Number Of Points
SENS:TRAC:POIN
312
Scope Realtime
SENS:TRAC:REAL OFF
Trigger Delay
TRIG:DEL
0 s
Trigger Level
TRIG:LEV
-30 dBm
Trigger Source
TRIG:SOUR INT
Trigger Slope
TRIG:SLOP POS
Trigger Holdoff
TRIG:HOLD
19.9 ms
Trigger Hysteresis
TRIG:HYST
3 dB