R&S NRP
Remote Control - Commands
1144.1400.12 6.101
E-3
Command
Preset and *RST value
Remark
TRIG[1..4|:ALL]:LEV
<NRf>
Depending on sensor.
TRIG[1..4|:ALL]:SLOP
POS | NEG
Depending on sensor.
TRIG[1..4|:ALL]:SOUR
IMM | HOLD | EXT
| INT | BUS
Depending on sensor.
UNIT[1..8]:POW DBM
Logarithmic result scaling.
UNIT[1..8]:POW:RAT DB
Logarithmic result scaling.
SYSTem:SENSor[1..4]:INFO? [<string>]
Returns information on a sensor.
<string>
is used to query a specific information item. If called
without parameters, the command returns all available information in the form of a list of strings
separated by commas. The permissible values for
<string>
depend on the sensor. For details refer to
the documentation of the corresponding sensor.
Value range:
depending on sensor
SYSTem:SENSor[1..4]:RESet
From the point of view of the R&S NRP base unit, the sensors are stand-alone measuring devices.
They communicate with the R&S NRP via a command set complying with SCPI.
SYST:SENS[1..4]:RES
prompts the base unit to send an
*RST
to the sensor in question.
Measurements in progress are interrupted.
SYSTem:SENSor[1..4]:TEST?
Performs a sensor test and returns a list of strings separated by commas. The contents of this test
protocol is sensor-specific. For its meaning, please refer to the sensor documentation.
SYSTem:SPEed NORMal | FAST
The data processing speed of the R&S NRP can be increased when
FAST
is selected. The display is
switched off and the measured values are no longer displayed since the continuous update of the
screen content requires computation time.
*RST value
:
NORM
SYSTem:STANdard:CATalog?
Returns a list of stored standards containing a series of strings separated by commas. The strings can
be used as parameters for the
SYSTem:STANdard:PRESet <string>
command.
*RST
value:
none