R&S NRP
Annex: Remote Control - Commands
1144.1400.12
Annex - I.3
E-2
Processing the results of two sensors
The R&S NRP offers measurement commands for processing the results of two sensors. The
calculated result can also be partly referenced to a stored reference value.
The measurement commands are obtained by adding one of the following keywords:
:DIFF
(difference)
:DIFF:REL
(difference with reference value)
:RAT
(ratio)
:RAT:REL
(ratio with reference value)
:SWR
(standing wave ratio)
:RLOS
(return loss)
:REFL
(reflection coefficient)
The list of parameters is modified for these commands since two sensors have to be entered in the
<source_list>
.
Example:
MEAS:RAT DEF,3,(@1),(@4)
A CALCulate block is used for processing the measured values. Each CALCulate block has two input
channels called the primary and secondary channel. In the above example, the primary channel is
assigned sensor 1 and the secondary channel sensor 4 via the
<source_list>
. The measurement
result is the ratio of the results from sensors 1 and 4.
Example:
MEAS:SWR DEF,2,(@1),(@2)
It is assumed that sensor 1 delivers the forward power of a wave and sensor 2 the reflected power of a
wave. The measurement result represents the SWR of the two power values:
( ) ( )
( ) ( )
1
/@
/
2
@
1
1
/@
/
2
@
1
−
+
=
P
Note that the unit is automatically set to percent (refer to Selecting the output unit for measured values).