Remote Control - Commands
1144.1400.12 6.87
E-3
Questionable Power Status Register
The CONDition register contains information as to whether the measured power values are
questionable.
The Questionable Power Status Register can be read by the following commands:
STATus:QUEStionable:POWer[:SUMMary]:CONDition?
STATus:QUEStionable:POWer[:SUMMary][:EVENt]?
Table 6-27: Meaning of bits used in the Questionable Power Status Register
Bit
No.
Meaning
0
Not used
1 Sensor
A
Power
The measurement data of sensor A is corrupt.
2
Sensor B Power
The measurement data of sensor B is corrupt.
3
Sensor C Power
The measurement data of sensor C is corrupt.
4
Sensor D Power
The measurement data of sensor D is corrupt.
5 Sensor
A
please
zero
The zero correction for sensor A is no longer correct and should be repeated.
6
Sensor B please zero
The zero correction for sensor B is no longer correct and should be repeated.
7
Sensor C please zero
The zero correction for sensor C is no longer correct and should be repeated.
8
Sensor D please zero
The zero correction for sensor D is no longer correct and should be repeated.
9 to
14
Not used
15
Bit 15 will never be used.
The sensor power bit is set when error -230, "Data corrupt or stale", or -231, "Data questionable",
occurs in the respective sensor.