R&S NRP
Remote Control - Commands
1144.1400.12 6.33
E-3
DISPlay[:WINDow][1..4]:METer|:ANALog:AUTo[?] ONCE | OFF
DISP:AUTO ONCE
automatically determines scaling for the analog display. The upper and the lower
limit value of the display is set as a function of the current measurement data. Only the
ONCE
parameter
is permissible; a query always supplies the value
OFF
.
*RST value
:
OFF
DISPlay[:WINDow][1..4]:METer|:ANALog:LOWer[?] <float_value>
Sets the lower limit value of the analog scale.
The value range depends on the current output unit of the measured value of the window.
Unit of measured value
W DBM
PCT DB
Value range for
<float_value>
1e-18..1e18 -150..210 1e-18..1e22 -200..200
Unit:
Æ
Table 6-3
on page
6.27
.
*RST value
:
-90 DBM
or
-120 DB
DISPlay[:WINDow][1..4]:METer|:ANALog:LOWer:POWer[?]
<float_value>
Sets the lower limit value of the analog scale. This value is used if the unit of the measurement result is
Watt, dBm or dBµV, i.e. the measurement result is a power.
Unit:
DBM | W | DBUV
Default unit:
DBM
Value range
:
see
DISPlay[:WINDow]:METer|:ANALog:LOWer[?]
*RST
value:
-90 DBM
DISPlay[:WINDow][1..4]:METer|:ANALog:LOWer:RATio[?]
<float_value>
Sets the lower limit value of the analog scale. This value is used if the unit of the measurement result is
dB, percent or the number 1, i.e. the measurement result is a power ratio.
Unit:
DB | DPCT | O
Default unit:
DB
Value range
:
see
DISPlay[:WINDow]:METer|:ANALog:LOWer[?]
*RST
value:
-120 DB