R&S NRP
Data acquisition and parameters (Sensor menu)
1144.1400.12 4.21
E-2
Filter dialog box
:
Auto Config…
–
( )
ä
ä
Fig. 4-18
Auto Filter dialog box
A | B | C | D
–
( )
ä
↔
ä
Tabs for selecting the sensor.
Normal –
( )
ä
Selects
the
autofilter
Normal
mode.
Remote Control
[SENSe[1..4]]:AVERage:COUNt:AUTO:TYPE[?] RESolution
Fixed Noise
–
( )
ä
Select the autofilter
Fixed Noise
mode.
Remote Control
[SENSe[1..4]]:AVERage:COUNt:AUTO:TYPE[?] RESolution
Noise Content
–
( )
ä
ä
Selects the proportion of intrinsic noise in the measured result. Specifi-
cally,
Noise Content
gives the permitted relative variation of the result
which may not be exceeded for 95% of the observation time.
Remote Control
[SENSe[1..4]]:AVERage:COUNt:AUTO:NSRatio[?] <float_value>
Max Settling Time
–
( )
ä
ä
Specifies an upper limit for the settling time in the
Fixed Noise
mode.
Limit exceeded
S/N
Remote Control
[SENSe[1..4]]:AVERage:COUNt:AUTO:MTIMe[?] <float_value>
Ref Timeslot
–
( )
ä
ä
Specifies the timeslot to which autofilter is applied when measure-
ments are being performed in the
Timeslot
mode. This parameter is not
available in the other measurement modes.
Remote Control
[SENSe[1..4]]:AVERage:COUNt:AUTO:SLOT[?] <int_value>