SCPI Error Messages
R&S NRP
1144.1400.12
9.4
E-2
Error code
Error text returned upon queue query
Explanations
-225
Out of memory
The device has insufficient memory to perform the required operation.
-226
Lists not same length
A list of offset values and frequencies has not the same length as an associated list of frequency and offset
values.
-230
Data corrupt or stale
The data are incomplete or invalid.
-240
Hardware error
The command cannot be executed because of a hardware problem in the device.
-241
Hardware missing
The command cannot be executed because of missing device hardware.
Device-Specific Errors
Device-specific error; causes bit 5 to be set in the ESR register
Error code
Error text returned upon queue query
Explanations
-300
Device-specific error
Generic device-dependent error that cannot be defined more precisely.
-310
System error
Indicates that some system error has occurred. Please contact the R&S service center.
-311
Memory error
An error was detected in the device’s memory.
-314
Save/recall memory lost
Indicates that the nonvolatile data saved by the *SAV? command has been lost.
-315
Configuration memory lost
Indicates that nonvolatile configuration data saved by the device has been lost.
-330
Self-test failed
The self-test could not be executed.
-341
Zeroing failed
The zeroing of a sensor (
CALibration:ZERO:AUTO ONCE
) has failed. The most frequent cause for this error is
a signal that is applied to the sensor input and prevents zeroing.
-350
Queue overflow
This code is entered into the queue in lieu of the code that caused the error. It indicates that an error occurred but
was not recorded. Five entries are accepted in the queue.
-363
Input buffer overrun
More commands were sent to the device than can be simultaneously executed or stored.