R&S NRP
Remote Control - Commands
1144.1400.12 6.103
E-3
TRIGger
The trigger commands are used to configure the timing conditions for the start of a measurement.
Note:
When the sensors are connected to the R&S NRP, they inform the device
about the SENSe and TRIGger commands supported and about the value
ranges for parameters. This applies to both numeric and text parameters.
For this reason, no fixed ranges are specified with these commands; the
ranges can be obtained from the respective sensor manual.
Table 6-34: Commands for setting the trigger system of a sensor
Command
Parameter
Unit
Remark
ABORt[1..4|:ALL]
No query
INITiate[1..4|:ALL]
:CONTinuous[?]
ON | OFF
-
:DISable[?]
ON | OFF
-
[:IMMediate]
-
-
No query
TRIGger:ALL
:SYNChronize[:STATe][?] ON | OFF
-
TRIGger[1..4|:ALL]
:ATRigger[:STATe][?]
ON | OFF
-
:COUNt[?]
<int_value>
-
:DELay[?]
<float_value>
S
:AUTO[?]
ON | OFF
-
:HOLDoff[?]
<float_value>
S
:HYSTeresis[?]
<float_value>
DB|PCT
[:IMMediate]
-
-
No query
:LEVel[?]
<float_value>
DBM|W
:SLOPe[?]
POSitive | NEGative
-
:SOURce[?]
BUS | EXTernal | HOLD |
IMMediate | INTernal
-
ABORt[1..4|:ALL]
Immediately sets the respective sensors to the IDLE state. Measurements in progress are interrupted. If
INIT:CONT ON
is set, a new measurement is immediately started since the trigger system is not
influenced.
INITiate[1..4|:ALL]:CONTinuous[?] ON | OFF
Selects either single-shot or continuous (free-running) measurement cycles. In the course of a
measurement cycle, a sensor passes through the IDLE, INITIATED, WAIT_FOR_TRG and
MEASURING states before it is again set to the IDLE (
INIT:CONT OFF
) or the INITIATED (
INIT:CONT
ON
) state. The settings are not changed when control is switched from manual to remote.
INIT:CONT
ON
is set when remote control is switched back to manual control.
INIT:CONT ON
Continuous measurements are performed. If a measurement is completed, the
respective sensors do not return to the IDLE state but are immediately set to
INITIATED and then to WAIT_FOR_TRG.
INIT:CONT OFF
A measurement cycle is only performed once. After completion, the sensors remain
in the IDLE state.
INIT:CONT OFF
has no effect when the sensor is already in the
IDLE state. A measurement in progress is completed.
*RST
-Wert:
(PRES)
and
SYST:PRES
:
ON
*RST
:
OFF