Data acquisition and parameters (Sensor menu)
R&S NRP
1144.1400.12 4.24
E-2
Trigger settings
The trigger system, which is required for the BurstAv, Timeslot and Scope measurement modes, is
configured in the
Trigger
dialog box and the
Trigger (Adv)
dialog box.
Sensor menu
:
Trigger… –
( )
ä
ä
Fig. 4-21
Trigger dialog box
A | B | C | D
–
( )
ä
↔
ä
Tabs for selecting the sensor.
Delay –
( )
ä
ä
Trigger delay. Setting a positive value delays the effect of the trigger
event until the set time has elapsed. Entering negative values allows
pretriggering (limited to a few ms, depending on the sensor).
The trigger delay is set in an editing field with direct entry of values
(see Chapter 3,
Dialog boxes and their controls
).
Tip
Entering a correct value for the trigger delay is absolutely essential for
accurate measurements in the Timeslot mode. For this reason, the
setting should always be checked in the Scope mode
(
T’slots,
p. 4.45).
Remote Control
TRIGger[1..4|:ALL]:DELay[?] <float_value>
Source –
( )
ä
ä
Drop-down list for the trigger source (external or internal) and trigger
slope (positive or negative).
Remote Control
TRIGger[1..4|:ALL]:SOURce[?]
BUS | EXTernal | HOLD | IMMediate | INTernal
Level –
( )
ä
ä
When this level is exceeded, a measurement is triggered (or when the
signal drops below this level in the case of a falling slope).
The trigger level is set in an editing field with direct entry of values (see
Chapter 3,
Dialog boxes and their controls
).
Remote Control
TRIGger[1..4|:ALL]:LEVel[?] <float_value>
Next… –
( )
ä
ä
Switches to the second page of the Trigger dialog box.
Sensor
Trigger…