![Rohde & Schwarz R&S NRP Operating Manual Download Page 179](http://html.mh-extra.com/html/rohde-and-schwarz/rands-nrp/rands-nrp_operating-manual_1477868179.webp)
Remote Control - Commands
R&S NRP
1144.1400.12 6.4
E-3
Common Commands to IEEE 488.1
The following messages are interface commands developed for the IEC/IEEE bus, which can be sent
when the ATN line is active (
Æ
section "
IEC/IEEE Bus Interface
", page
6.124
).
DCL
- Device CLear
All commands and messages in progress are aborted. The TRIGger system goes to the IDLE state,
input and output queues are cleared, and the parser is reset and waits for the beginning of a new
command. When INITiate:CONT ON is set, a new measurement is started. This command makes the
device ready to receive commands.
GTL
- Go To Local
The device is set to local mode and can again be operated manually.
LLO
- Local Lock Out
Deactivates the front-panel keypad with the exception of the power key
( Ø )
. In the LLO state, it is
possible to return to the LOCAL mode only by remote control (
GTL
command) or switching the
R&S NRP off and on again.
PPC
- Parallel Poll Configure
Sets the device to the "Parallel Poll Addressed to Configure State" (PACS).
PPD
- Parallel Poll Disable
In the PACS, the parallel poll function is deactivated with
PPD.
PPE
- Parallel Poll Enable
In the PACS, the parallel poll function is activated with
PPE
and the bit position and the sense bit are
set.
PPU
- Parallel Poll Unconfigure
This command deactivates the parallel poll function without previous addressing.
SDC
- Selected Device Clear
Same as
DCL