Remote Control - Commands
R&S NRP
1144.1400.12 6.50
E-3
SENSe (Sensor Commands)
The commands of the SENSe group configure the power sensors with the exception of the trigger
system. This includes the measurement modes (ContAv, Burst, Timeslot or Scope), the offset
corrections and the filter settings.
Note:
When the sensors are connected to the R&S NRP, they inform the device
about the SENSe and TRIGger commands supported and about the value
ranges for parameters. This applies to both numeric and text parameters.
For this reason, no fixed ranges are specified with these commands; the
ranges can be obtained from the respective sensor manual.
Table 6-10: Commands of the SENSe system
Command
Parameter
Unit
Remark
[SENSe[1..4]]
:AVERage
[:STATe][?]
ON|OFF
-
:COUNt[?]
<int_value>
-
:AUTO[?]
ON|OFF|ONCE
-
:MTIMe[?]
<float_value>
S
:RESolution[?]
1|2|3|4
-
:SLOT
<int_value>
-
:NSRatio[?]
<float_value>
DB|PCT
:TYPE[?]
RESolution|NSR
-
:RESet
-
-
no query
[:STATe][?]
ON|OFF
-
:TCONtrol[?]
MOVing|REPeat
-
:CORRection
:OFFSet[?]
<float_value>
DB|PCT
:STATe[?]
ON|OFF
-
:DCYCle
[:INPut][:MAGNitude][?]
<float_value>
PCT
:STATe[?]
ON|OFF
-
:FDOFfset
[:INPut][:MAGNitude]?
-
DB
:FDOTable
[:SELect][?]
<string>
-
:STATe[?]
ON|OFF
-
:SPDevice:STATe[?]
ON|OFF
-
:FREQuency[:CW|:FIXed][?]
<float_value>
HZ
:FUNCtion[:ON][?]
<string>
-
[:POWer]
[:AVG]
:APERture[?]
<float_value>
S
:BUFFer:SIZE[?]
<int_value>
-
:BUFFer:STATe[?]
ON|OFF
-
:RANGe[?]
0|1|2
-
:AUTO[?]
ON|OFF
-
:AUTO:CLEVel[?]
<float_value>
DB|PCT
:SMOothing:STATe[?]
ON|OFF
-
:TGATe[1..4]
:OFFSet:TIME[?]
<float_value>
S
:TIME[?]
<float_value>
S
:TGATe:SELect[?]
1|2|3|4
:TSLot
:COUNt[?]
<int_value>
-
:WIDTh[?]
<float_value>
S