![Rohde & Schwarz R&S NRP Operating Manual Download Page 269](http://html.mh-extra.com/html/rohde-and-schwarz/rands-nrp/rands-nrp_operating-manual_1477868269.webp)
Remote Control - Commands
R&S NRP
1144.1400.12 6.94
E-3
SYSTem:BEEPer:NOTify:KEY[?] ON | OFF
Enables (
ON
) or disables (
OFF
) acoustic key clicks, respectively.
*RST
-Wert:
OFF
SYSTem:BEEPer:STOP
Stops the output of an audible signal.
SYSTem:BEEPer:TIME <NRf>
Sets the length in milliseconds of the audible signal output with
SYS:BEEP.
Value range
:
1..60
Unit:
S
*RST value
:
This setting is not changed by
*RST
.
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <NR1>
Sets the address with which the R&S NRP can be addressed via the IEC/IEEE bus. The address is
factory-set to 20 and is not changed by a reset.
Value range:
0..30
*RST
value:
This setting is not changed by
*RST
.
SYSTem:DATE[?] <year>,<month>,<day>
Parameter
Value range
<year>
2000 ... 2100
<month>
1
(January)
... 12
(December)
<day>
1 ... 31
Sets the date. Since the R&S NRP has no battery-backed clock, the date has to be reset if the clock
was not set after the last power-on of the R&S NRP.
*RST value
:
This setting is not changed by
*RST
.