Displaying data in the Scope mode
R&S NRP
1144.1400.12 4.46
E-2
Marks the set timeslot including the exclude start and end times.
Average
Average power for the set timeslot. The displayed value is obtained by
averaging the measurement points between the inner vertical lines.
Measurement accuracy can be increased if the Timeslot mode is se-
lected for the actual measurement.
Slot# –
( )
ä
↔
ä
Selection of the marked timeslot.
For this purpose, an average power calculated from the measurement
points is displayed.
Excluded from End –
( )
ä
… from Start
–
( )
ä
Setting of the exclude times.
These parameters can also be set in the
Mode
dialog box.
(
Measuring the power of TDMA signals, p. 4.13).
Tip for power measurement in timeslots
1. Enter TDMA standard parameters in the Mode menu (Timeslot).
2. Switch the sensor to the Scope mode.
3. Make the horizontal and vertical display settings on the
Trace
page.
4. Blend in the timeslot structure on the
T'slots
page.
5. Open the Trigger dialog box and set the delay so that the timeslot structure coincides with the trace.
6. Change the exclude times if necessary.
Gates
The R&S NRP provides four separately configurable time windows ("gates") for signal evaluation. The
start and end of each gate can be set on the
Gates
page with the aid of markers.
Measurements in the time windows can be made by using the
Meas
page (
Meas, page 4.47), or by
switching to the Timegate mode (
Measuring power in gates, p. 4.14).
Scope mode display
…Trace,
Gates… –
( )
ä
↔
ä
Fig. 4-40
Scope mode display – Gates page