Remote Control - Commands
R&S NRP
1144.1400.12 6.122
E-3
Compatibility Information
This section contains information regarding the compatibility of the R&S NRP remote control commands
with those of the E4418B and E4419B power meters from Agilent.
The R&S NRP is largely downward-compatible to the above-mentioned devices but offers additional
capabilities.
Table 6-36: Compatibility information about remote control commands
E4418B/E4419B command
Compatibility information
*DDT
Is not implemented in the R&S NRP.
CONFigure?
Returns the current configuration. The R&S NRP returns the last setting
made with
CONFigure,
as specified by the SCPI standard.
CALCulate:GAIN
Is not implemented in the R&S NRP.
CALCulate:PHOLd
Is implemented in the R&S NRP and has the same effect as
CALC:EXTR:RES
.
CALibration
:ECONtrol
:RCALibration
:RFACtor
Is not implemented in the R&S NRP.
DISPlay:ENABle[?]
Is implmented in the R&S
NRP and has the same effect as
DISPlay:ILLUmination
.
DISPlay:SCREen:FORMat
[:NUMeric[1]|2]
Is not implemented in the R&S NRP.
INITiate
:CONTinuous:SEQUence[1]|2
:IMMediate:SEQUence[1]|2
The
:SEQUence
node is not implemented in the R&S NRP.
MEMory:TABLe:GAIN
Agilent specifies the values without a unit; the unit
PCT
(percent) is
implicitly assumed. The units
DB
(dB),
DPCT
(
∆
%) and
O
(unit 1) can
be specified in the R&S NRP. The default unit is
DB
.
OUTPut:TRIGger:STATe
Is not implemented in the R&S NRP.
SENSe
:BANDwidth
:AVERage:COUNt:AUTO:SDETect
:DETector:FUNCtion
:MRATe
:SPEed
:V2P
Is not implemented in the R&S NRP.
SENSe[1|2]:CORRection
:CSET[1]|CSET2[:SELect]
and
SENSe[1|2]:CORRection
:CSET[1]|CSET2:STATe
The R&S
NRP does not recognize any tables with frequency-
dependent calibration values since the sensors have to be calibrated
only once in production. For this reason, there is no compliance with
CSET1.
Tables with frequency-dependent offset correction values
can be defined and addressed via :
CSET2
(Agilent and R&S NRP) or
.
:FDOTable
(R&S NRP only).
SENSe[1|2]:CORRection:GAIN[1]?
Has the same effect as