Remote Control - Commands
R&S NRP
1144.1400.12 6.108
E-3
UNIT
Table 6-35: Commands of the UNIT system
Command
Parameter
Unit
Remark
UNIT[1..8]
:POWer[?]
DBM | W
-
:RATio[?]
DB | PCT
-
UNIT[1..8]:POWer[?] DBM | W | DBUV
Selects the output unit for the measured power values. This setting also determines the unit for the
parameters of the following commands:
•
CALC:REL
•
CALC:LIM:UPP
•
CALC:LIM:LOW
•
DISP:MET:UPP
•
DISP:MET:LOW
•
OUTP:REC:LIM:LOW
•
OUTP:REC:LIM:UPP
*RST
value:
DBM
UNIT[1..8]:POWer:RATio[?] DB | DPCT | O
Selects the output unit for the measured power ratio values. This setting also determines the unit for the
parameters of the following commands:
•
CALC:REL
•
CALC:LIM:UPP
•
CALC:LIM:LOW
•
DISP:MET:UPP
•
DISP:MET:LOW
•
OUTP:REC:LIM:LOW
•
OUTP:REC:LIM:UPP
*RST
value:
DB
Error messages
: If the compute function of the associated calculate block is
:SWR
,
:RLOSs
or
:REFLection
, then the unit cannot be changed and error message 26,"State not supported", is output.