R&S NRP
Remote Control - Commands
1144.1400.12 6.99
E-3
Command
Preset and *RST value
Remark
DISP[1..4]:TRAC:UPP 30
DBM
Upper limit of power axis in Scope
mode
DISP[1..4]:TSL 1
First timeslot
DISP[1..4]:UPD NORM
Frequent display updates.
FORM
ASC
Measured data in ASCII format.
FORM:BORD
NORM
"Big-Endian" byte sequence of
measured data, if FORM REAL.
MEM:...
-
The settings in the MEMory command
system will not be affected.
OUTP:REC1:FEED "CALC1"
Analog outputs are fed by the first
calculate block.
OUTP:REC2:FEED "CALC2"
Analog outputs are fed by the second
calculate block.
OUTP:REC[1..2]:LIM:LOW -20
DBM
or
0 DB
0 V equivalent at analog output.
OUTP:REC[1..2]:LIM:UPP 20
DBM
or
10 DB
3 V equivalent at analog output.
OUTP:REC1:STAT ON
Analog output active.
OUTP:REC2:STAT OFF
Trigger input active.
OUTP:ROSC
OFF
Test generator is switched off.
OUTP:TTL:ACT LOW
OUTP:TTL:LVOLT is applied to the
TTL output if a limit is violated.
OUTP:TTL:FAIL HIGH
OUTP:TTL:HVOLT is applied to the
TTL output if a limit is violated.
OUTP:TTL:FEED "CALC1:LIM"
The TTL output is fed by the limit
monitoring function of calculate block
1.
OUTP:TTL:HVOLT 3.3
V
High voltage at TTL output
OUTP:TTL:LVOLT 0.0
V
Low voltage at TTL output
SENS[1..4]:AVER
ON | OFF
Depending on sensor.
SENS[1..4]:AVER:COUN <NR1>
Depending on sensor.
SENS[1..4]:AVER:COUN:AUTO OFF
Depending on sensor.
SENS[1..4]:AVER:COUN:AUTO:MTIMe <NRf>
Depending on sensor.
SENS[1..4]:AVER:COUN:AUTO:RES <NRf>
Depending on sensor.
SENS[1..4]:AVER:COUN:AUTO:SLOT <NR1>
Depending on sensor.
SENS[1..4]:AVER:COUN:AUTO:NSR <NRf>
Depending on sensor.
SENS[1..4]:AVER:COUN:AUTO:TYPE
RES | NSR
Depending on sensor.
SENS[1..4]:AVER:TCON MOV
REP
(PRES)
and
SYST:PRES
.
*RST
.
SENS[1..4]:CORR:DCYC <NRf>
Depending on sensor.
SENS[1..4]:CORR:DCYC:STAT ON
|OFF
Depending on sensor.
SENS[1..4]:CORR:FDOT -
No change.
SENS[1..4]:CORR:FDOT:STAT OFF
SENS[1..4]:CORR:OFFS <NRf>
Depending on sensor.
SENS[1..4]:CORR:OFFS:STAT
ON | OFF
Depending on sensor.
SENS[1..4]:FREQ -
No change.
SENS[1..4]:FUNC "POW:AVG"|
"POW:BURS:AVG"|
"POW:TSL:AVG"|
"XTIM:POW"
Depending on sensor.