R&S NRP
Remote Control - Commands
1144.1400.12 6.47
E-3
OUTPut:RECorder[1..2]:STATe[?] ON | OFF
Switches the analog outputs (OUT1 and OUT2/TRIG on rear panel)
ON
or
OFF
. The magnitude of the
output voltage (0 V to 3.3 V) depends on the measured power of the assigned calculate block
(
OUTP:REC:FEED
) and the configuration of the characteristic (
OUTP:REC:LIM
). As in manual
operation, this setting is coupled to other operating modes of the outputs. Only one mode can be active
at any one time:
Connector
Analog output
(
OUTP:REC
)
TTL output
(
OUTP:TTL
)
Off
Ext. Trigger In
OUT1
x x x -
OUT2/TRIG
x - - x
*RST
value:
output 1:
ON
output 2:
OFF
OUTPut:ROSCillator[:STATe][?] ON | OFF
The command switches the optional RF generator (
ON
) or (
OFF
).
*RST value
:
OFF
OUTPut:ROSCillator:CALibration:READ? <NR1>
Reads the setting of one of the internal electronic potentiometers of the test generator (option
R&S NRP-B1). The number of the potentiometer (1 to 4) is indicated with the parameter <NR1>. The
response is an 8-bit integer with a sign (-128 to 127).
*RST
value:
none
OUTPut:ROSCillator:CALibration:WRIte <NR1>,<NR1>
Sets the test generator (option R&S NRP-B1). Refer to the following table for the meaning of the two
parameters:
1st parameter <NR1>
01: Generator on/off (2nd parameter 0: on / 1: off)
02: Power stage (2nd parameter 0: -20 dBm, 1: 0 dBm)
03: LED state (2nd parameter 0: off, 1: on, 0x02: blinking)
04: Start of calibration
05: Setting of potentiometer 1 (2nd parameter –128 to 127)
06: Setting of potentiometer 2 (2nd parameter –128 to 127)
07: Setting of potentiometer 3 (2nd parameter –128 to 127)
08: Setting of potentiometer 4 (2nd parameter –128 to 127)
09: Confirm calibration
10: Abort calibration
2nd parameter <NR1>
Data if 1st parameter 01, 02, 03, 05, 06, 07 or 08.
0 if 1st parameter 04, 09 or 10.
*RST
value:
none