Remote Control - Fundamentals R&S
NRP
1144.1400.12 5.2
E-3
Measurement Modes
To simplify operation in the manual mode, preset values are used in some cases for the trigger system
(TRIGger command) and the timing of the measurement (
SENSe[1..4]:TIMing:EXCLude:STARt
and :STOP
) when the measurement mode (ContAv, Burst, Timeslot, Scope)
is selected (
Æ
Table 5-2
). This is not the case in the remote control mode. Here the trigger system and
the timing can be configured by the user as required.
Calculate Blocks
The R&S NRP manages 8 internal calculate blocks. Each calculate block has two input channels
(primary and secondary channel) to which one of up to 4 power sensors can be assigned. The channels
can be processed together by means of a selectable function. The result is available at the output of the
calculate block. In the manual mode, the measurement results of the first four calculate blocks
(
Æ
"
CALCulate System
", page
6.10
) are shown in windows 1 to 4, while 8 calculate blocks are available
in the remote control mode.