R&S NRP
Remote Control - Commands
1144.1400.12 6.55
E-3
[SENSe[1..4]]:CORRection:FDOTable[:SELect][?] <string>
The name of the current frequency-dependent offset table can be set or queried by the
[SENSe[1..4]]:CORRection:FDOTable[:SELect][?]
command. The R&S NRP can manage up
to ten such tables (MEMory commands). The frequency-dependent offset tables consist of two
columns; the first containing the frequencies, the second the associated correction values. When the
R&S
NRP is informed about the signal frequency with command
[SENSe[1..4]]:FREQuency[:CW|:FIXed] <float_value>
, an appropriate correction factor is
determined by means of the table. If the table was activated with
[SENSe[1..4]]:CORRection:FDOTable:STATe[?] ON
, the measurement result obtained by the
sensor is multiplied by this factor.
*RST value
:
Settings are not changed in case of
*RST
.
Error messages:
-256 "
File name not found
": A table named
<name>
is not known
.
[SENSe[1..4]]:CORRection:FDOTable:STATe[?] ON | OFF
This command activates and deactivates the current frequency-dependent offset table. The current
table is selected with the [
SENSe[1..4]]:CORRection:FDOTable[:SELect]
command. Only one
table can be active at a time.
*RST value
:
OFF
[SENSe[1..4]]:CORRection:SPDevice:STATe[?] ON | OFF
Instructs the sensor to perform a measured-value correction by means of the stored s-parameter
device.
*RST value
:
depending on sensor
[SENSe[1..4]]:FREQuency[:CW|:FIXed][?] <float_value>
This command informs the R&S NRP about the frequency of the power to be measured since this
frequency is not automatically determined. The frequency is used to determine a frequency-dependent
correction factor for the measurement results.
Unit:
HZ
Value range:
depending on sensor
*RST value
:
depending on sensor