
Remote Control - Commands
R&S NRP
1144.1400.12 6.88
E-3
Questionable Window Status Register
The CONDition register contains information as to whether the displayed data or the power calculated
by the calculate blocks is questionable.
The Questionable Window Status Register can be read by the following commands:
STATus:QUEStionable:WINDow[:SUMMary]:CONDition?
STATus:QUEStionable:WINDow[:SUMMary][:EVENt]?
Table 6-28: Meaning of bits used in the Questionable Window Status Register
Bit
No.
Meaning
0
Not used
1 Window
1
Power
The measured values returned by Calculate block 1 are corrupt.
2 Window
2
Power
The measured values returned by Calculate block 2 are corrupt.
3 Window
3
Power
The measured values returned by Calculate block 3 are corrupt.
4 Window
4
Power
The measured values returned by Calculate block 4 are corrupt.
5 Window
5
Power
The measured values returned by Calculate block 5 are corrupt.
6 Window
6
Power
The measured values returned by Calculate block 6 are corrupt.
7 Window
7
Power
The measured values returned by Calculate block 7 are corrupt.
8 Window
8
Power
The measured values returned by Calculate block 8 are corrupt.
9 to
14
Not used
15
Bit 15 will never be used.
A bit is set if the error -231, "Data questionable", occurs in the respective calculate block.