R&S NRP
Displaying measurement results (Windows menu)
1144.1400.12 4.31
E-2
Setting the analog scale
The analog scale in the combined digital/analog display (
D/A
) can be set with the
Analog Meter
softkey.
Windows menu
:
Analog –
( )
Meter…
ä
Fig. 4-27
Meter Scaling dialog box
Auto Once
–
( )
ä
ä
Automatically selects the appropriate scaling so that the instantaneous
measured value is in the middle of the scale.
Remote Control
DISPlay[:WINDow][1..4]:METer|:ANALog:AUTo[?] ONCE | OFF
Left –
( )
ä
ä
Enables input of the left end-of-scale value.
Remote Control
DISPlay[:WINDow][1..4]:METer|:ANALog:LOWer[?] <float_value>
Right –
( )
ä
ä
Enables input of the right end-of-scale value.
Remote Control
DISPlay[:WINDow][1..4]:METer|:ANALog:UPPer[?] <float_value>
Scrolling between timeslots
If the timeslot mode has been selected for a sensor, you can use the
Time Slot
softkey in the Windows menu to select the timeslot whose
measured values you want to display.
Fig. 4-28
Windows menu, softkey for scrolling between timeslots
Windows menu
:
Time Slot
–
( )
[1]
ä
↔
ä
Specifies which of the timeslots on which a simultaneous measurement
is being performed are displayed in the active window.
If you want to display further timeslots simultaneously, a separate win-
dow must be opened and configured for each additional timeslot.
Remote Control
DISPlay[:WINDow][1..4]:TSLot[?] <int_value>
Windows
Scale
Windows
Time Slot