R&S NRP
Remote Control - Commands
1144.1400.12 6.107
E-3
TRIGger[1..4|:ALL]:SOURce[?]
BUS | EXTernal | HOLD | IMMediate | INTernal
Sets the trigger signal source for the WAIT_FOR_TRG state.
BUS
The trigger event is initiated by
TRIG:IMM
or
*TRG
. In this case, the other trigger setting
are meaningless.
EXTernal
Triggering is performed with an external signal applied to the trigger connector. The
TRIG:SLOP
command determines whether the rising or the falling edge of the signal is
to be used for triggering. Waiting for a trigger event can be skipped by
TRIG:IMM
.
IMMediate
The sensor does not remain in the WAIT_FOR_TRG state but immediately changes to
the MEASURING state.
HOLD
A measurement can only be triggered when the command
TRIG:IMM
is executed.
INTernal
The sensor determines the trigger time by means of the signal to be measured. When
this signal exceeds (
TRIG:SLOP POS
) or fall short of (
TRIG:SLOP NEG
) the power set
by
TRIG:LEV
, the measurement is started after the time set by
TRIG:DEL.
Similar to
TRIG:SOUR EXT,
waiting for a trigger event can also be skipped by
TRIG:IMM
.
*RST value
:
depending on sensor