R&S NRP
Remote Control - Commands
1144.1400.12 6.127
E-3
Interface Messages
Interface messages are transmitted to the device on data lines, where the ATN is set to active LOW.
They are used for communication between the controller and the R&S NRP.
Common Commands
Common commands are in the code range 0x10 to 0x1F. They affect all devices on the bus without any
addressing being required.
Table 6-38: Common commands
Command
Effect on device
DCL (Device Clear)
Interrupts processing of received commands and sets the command processing
software to a defined initial state. This command does not change the device setting.
IFC (Interface Clear)
Resets the interfaces to their initial condition.
LLO (Local Lockout)
Manual switchover to LOCAL is disabled.
SPE (Serial Poll Enable)
Ready for serial poll.
SPD (Serial Poll Disable)
End of serial poll.
PPU (Parallel Poll Unconfigure)
End of parallel poll query status.
Addressed Commands
Addressed commands are in the code range 0x00 to 0x0F hex. They only affect devices addressed as
a listener.
Table 6-39: Addressed commands
Command
Effect on device
SDC (Selected DeviceClear)
Interrupts processing of received commands and sets the command processing
software to a defined initial state. This command does not change the device setting.
GTL (Go to Local)
Change to local mode (manual control).
PPC (Parallel Poll Configure)
Configures device for parallel poll.