Displaying data in the Scope mode
R&S NRP
1144.1400.12 4.44
E-2
Setting parameters
All parameters relevant to the trace display can be directly set in the expanded window. The parameters
are spread across four pages marked by tabs.
Symbols
Tabs for selecting the page.
Marks the trigger level (visible at the left- and right-hand edges). If the
Trigger dialog box is open, an unbroken line appears.
Time of trigger event.
Time at which the trigger delay has elapsed (delayed trigger). In the
Timeslot mode, this time should coincide with the beginning of the first
timeslot.
Tip
If you use the up/down cursor keys for making changes to parameters
(
Entry fields on page 3.8), you can observe the effects immediately
and make adjustments interactively. Coarse or fine adjustments are
possible, depending on the decimal place in which you "scroll".
Trace
The parameters on the Trace page are used to determine the position and the size of the screen win-
dow. The number of measurement points must be specified in the Mode menu (
Switching the sensor
to the Scope mode, p. 4.41).
Scope mode display
T’slots,
Trace… –
( )
ä
↔
ä
Fig. 4-38
Scope mode display – Trace page
Start –
( )
ä
Start of the measurement window (left-hand screen edge) relative to
the delayed trigger.
The value can be negative so that signal components before the trigger
event can be shown (limited to a few ms, depending on the sensor).
Remote control
[SENSe[1..4]]:TRACe:OFFSet:TIME[?] <float_value>