Clarus SQ 8 MS Hardware Guide
162
Chromatography Related
Problem
Probable Cause
Solution
Inconsistent retention
time.
Injector septum leak.
Replace the septum.
Carrier gas manifold leak.
Locate and fix the leak.
Rising Total Ion
Chromatogram baseline.
Column bleed.
Disconnect the column from
the mass spectrometer and
condition the column.
Vacuum leak.
Locate and fix the leak.
Discreet high intensity
contaminant peaks.
Column bleed.
Disconnect the column from
the mass spectrometer and
condition the column.
Injector septum bleed.
Replace the septum and/or
glass liner.
Tailing peaks (sloping on
RHS).
Improperly installed column.
Check the column and
reinstall if necessary.
Injector too cool.
Raise the injector
temperature.
Interface temperature too cool. Raise the interface
temperature.
Inadequate carrier gas flow.
Set proper flow.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......