Clarus SQ 8 MS Hardware Guide
120
Figure 34 Sliding Lock attachment on the connector
.
12.
Remove the RF Generator Cable connector. See the following figure.
Figure 35 Removing the RF Generator cable.
NOTE: The metal tube attached to the reference gas vacuum hose has a ferrule on it that may
slip off when the nut is removed from the reference gas valve fitting. Be careful to
avoid dropping the ferrule inside the instrument.
Press Down to
Release the Lock