Maintenance
125
Replacing the Electron Multiplier
The electron multiplier lifetime is dependent on the number of ions detected. You can
increase its life by correctly using the solvent delay settings and minimizing air leaks.
To replace an electron multiplier:
1.
Remove the ion optics assembly. See Removing and Returning the Ion Optics
Assembly on page 114
2.
Release the electron multiplier by pressing the retainer clip then pull the
electron multiplier from the assembly. See the following figure.
Figure 40 Removing the electron multiplier.
Quadrupole
Assembly
Electron Multiplier
Electron Multiplier
Pulled from the assembly
Electron Multiplier
Retainer Clip