Clarus SQ 8 MS Hardware Guide
156
Problem
Probable Cause
Solution
Solvent tailing.
Inadequate splitter flow.
Increase the splitter flow.
Column not properly
installed in the injector.
Reinstall the column in the
injector.
Loss of high end compounds.
Temperature setting too
low on the injector,
column oven or transfer
line.
Increase the injector, column
oven or transfer line
temperature to allow the less
volatile compounds of the
sample to reach the mass
spectrometer.
Peaks at masses 28 (nitrogen)
and 32 AMU (oxygen) are in a
ratio of < 4 to 1, and the peak at
mass 28 is larger than the peak
at mass 18.
Leaks in or around
vacuum or column
fittings.
Tighten the fittings and
connection points to the high
vacuum system.
Set the injector split flow to
50:1.
Peaks at masses 14 and 16 are
larger than the peak at mass 28.
Leaks or improper tuning.
Tighten the fittings and
connection points to the high
vacuum system.
Set the injector split flow to
50:1.
Run AutoTune.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......