Troubleshooting
167
Problem
Probable Cause
Solution
Contamination.
Locate the contamination and
eliminate it.
Co-eluting components.
Change your sample
preparation or
chromatography.
Incorrect column alignment.
Reinstall the column.
Section of a mass range
missing from a
spectrum.
Corruption of data file.
Reacquire data.
Scanning too fast.
Reduce the rate.
Hard disk has too much
fragmentation.
Defrag the hard drive.
Hard disk full.
Remove unnecessary files.
Molecular Ion too
weak.
Source temperature too high.
Reduce the source
temperature.