Clarus SQ 8 MS Hardware Guide
110
Figure 28 Removing the source parts for cleaning.
Cleaning
NOTE: You can do the following cleaning method of aluminum oxide paste to polish the flat
surfaces of the parts. For either cleaning method, the final step you must sonicate the
parts in an ultrasonic bath of methanol for at least five minutes. Dry off the parts using
lint-free tissue and/or clean compressed Nitrogen gas to prevent solvents from drying
on these parts and leaving a residue
1.
Mix together aluminum oxide and de-ionized water and a few drops of
methanol to make a watery paste.
2.
Dip a wooden-stick cotton swab or a foam-pad swab in the solution and clean
the darkened areas on the source. Work quickly to prevent the mixture from
drying on the surface.
3.
Place the cleaned components in de-ionized water prior to rinsing to prevent drying.
Rinsing
1.
Add 50 mL of methanol to a 100 mL beaker, insert the source parts, and
sonicate them in an ultrasonic bath for at least ten minutes.
2.
Carefully drain the methanol.
3.
Add 50 mL of acetone to the 100 mL beaker, insert the source parts, and
sonicate in an ultrasonic bath for ten minutes.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......