Troubleshooting
175
Message Title
Icon
Message
Action
Backing Pump is
on
The system is not in an operating
state. A vacuum leak or a vacuum
gauge failure could have occurred.
If the transfer line or the source
temperatures are above 100C, please
wait until they have cooled before
pressing OK.
Pressing the OK button will turn off
the backing pump.
If the transfer line or the
source temperatures are
above 100C, please wait
until they have cooled
before pressing
OK
.
Pressing the
OK
button
will turn off the backing
pump.
Problem with
Vacuum
The backing pump could not reach
the necessary vacuum level to start
the turbo pump.
There could be a problem with a
vacuum leak, the backing pump or
the vacuum gauge.
The backing pump will be turned off.
Make sure that the vent valve is
closed before restarting the backing
pump.
Press
OK
to close the
message.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......