Hydrogen
Ensure that all hydrogen lines and connections are leak-free.
When using a hydrogen tank, install an in-line hydrogen snubber
(part number 00090038) between the tank regulator and the
delivery tubing.
Ventilation
Always provide adequate ventilation. When analyzing hazardous
compounds such as pesticides, it may be necessary to arrange to
vent the mass spectrometer effluent from the forepump into a
fume hood. To prevent contamination if a fume hood is
unavailable, an oil separation filter should be installed at the
outlet of the forepump vented to a fume hood or an oil mist
separator (Alcatel 68316) with a charcoal trap (Koby KA1). An
acceptable alternative is to attach a ½ inch Tygon tube and vent
to a hood. Pump oil vapor is considered toxic and must be vented
properly.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......