Clarus SQ 8 MS Hardware Guide
92
10.
Slide the column into the Transfer Line tube until the septum is aligned at a
distance exactly 10 mm away from the end of the column nut. See
Figure 16
.
Figure 16 Measuring from the outlet of the column to the front of the septum.
11.
Using the 5 mm and 1/4 inch crescent wrenches, tighten the column nut until
the ferrule is crimped onto the column. See
Figure 17
.
Figure 17 Tightening the column nut.
metric
10
20
30
40
50
60
70
8
Slide the column back until the
septum is 10 mm from nut.
1/16-inch Column Nut
Septum
Column