About the Clarus GC/MS System
59
CAUTION
ATTENTION
The AC line cord for the rotary vane pump must be plugged into the
designated receptacle on the back of the Clarus MS. The pump is
controlled by the TurboMass software.
Connecting the vacuum hose to the exhaust connection of the rotary
pump will severely contaminate the Clarus MS.
Le cordon de ligne AC pour la pompe rotative doit être branché dans la
prise désignée à l'arrière du Clarus MS. La pompe est contrôlée par le
logiciel TurboMass.
La connexion du tuyau d'aspiration au raccord d'échappement de la
pompe rotative contaminera sévèrement le MS de Clarus.
Instrument Power Rotary Pump Outlet
120 VAC
120 VAC ~ 50/60 Hz MAX POWER 828 VA
230 VAC
230 VAC ~ 50/60 Hz MAX POWER 690 VA
Vacuum Pump Options
The Clarus SQ 8 MS offers two different vacuum pump capacities.
Turbomolecular pumps are high-speed turbines which transport the sample and
carrier gas molecules away from the mass spectrometer.
•
Clarus SQ 8 S - The 75 L/sec turbomolecular pump supports Electron
Ionization operation (EI) and has optional water cooling.
•
Clarus SQ 8 T – All of the functions and options of the SQ 8 S with a 255
L/sec turbomolecular pump for higher column flow rates, pump-down time
under three minutes, and lower detection limits
•
Clarus SQ 8 C - All of the functions and options of the SQ 8 T with positive
and negative Chemical Ionization (CI) operation.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......