Troubleshooting
165
Problem
Probable Cause
Solution
Poor reproducibility.
Dirty source.
Remove and clean the source.
Defective injector liner.
Replace injector liner.
Defective syringe.
Replace syringe.
Old or damaged filament.
Examine and replace
filament.
Poor tuning.
Retune.
Poor calibration.
Recalibrate.
Air leak.
Locate and fix.
Active sites in column/liner.
Replace column/liner.
Intermittent source heater
failure.
Call a PerkinElmer service
engineer.
Poor S/N on test standards. See low sensitivity causes
above.
Incorrect GC/MS method
Use the correct method.
Accidental split injection.
Set the proper split.
Detector voltage set too low.
Increase multiplier voltage.
Column flow rate too high.
Reset the column flow rate.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......