Maintenance
143
8.
Tuning may be optimized on the m/z 69, 219, 414, and 652 ions of the
heptacosa reference gas.
9.
Click
Press for Standby
to turn off Operate, followed by the CI gas.
You are now ready to run your CI+ analysis.
Adjust the Reagent Gas for CI-
1.
Open the CI gas inlet by selecting CI Gas from the Gas menu. Wait at least 10
seconds before clicking
Press for Operate
.
Parameter:
CI
−
value and comments:
Electron Energy
30 to 70 eV
(This parameter should be optimized.)
Emission
200 to 300
µ
A is acceptable
(Emission measures the real emission current, i.e.
the source current from the source block, there is no
separate measurement of source current in CI.)
Lens 1 and 2
The tuning of these lenses may be
different from the optimum values set for
EI, since the source pressure is much
higher in CI.
Multiplier
1300V to 1600V
Source temperature
150
°
C is standard. Higher temperatures
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......