Clarus SQ 8 MS Hardware Guide
160
Problem
Probable Cause
Solution
Piece of column broken off in
the ion chamber.
Remove the inner source,
check for and remove piece of
column.
Source filament is bent.
Check and replace filament if
necessary.
RF generator malfunction.
Call a PerkinElmer service
engineer.
Analyzer drive electronics
malfunction.
Call a PerkinElmer service
engineer.
Total Ion Chromatogram
too high.
Dirty source.
Clean the source.
Contamination from poor
handling technique.
Set source and transfer line to
250 ºC and maintain this
temperature overnight.
Stationary phase of column
de-polymerizing (bleeding).
Change column.
Air leak.
Find the leak and fix it.
Poor quality carrier gas.
Replace the carrier gas tank.
Carrier gas filter is ineffective
and needs replacing.
Replace the carrier gas filter.
No noise on mass
chromatogram.
Detector multiplier voltage too
low.
Increase the multiplier
voltage.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......