Warnings and Safety Information
Reagent Gases:
Ammonia
Methane
Isobutane
Minimum purity of 99.95%.
The gas delivery pressure required is 15 psi (103 kPa) to the
bulkhead fitting (1/8 in. Swagelok) on the mass spectrometer.
If ammonia is used for chemical ionization, all fittings and tubing
must be stainless steel to avoid corrosion. A single-stage regulator is
required for ammonia, rated for corrosive service. Also, the
forepump must be vented to a fume hood or trap.
A high-purity, stainless-steel diaphragm, two-stage regulator is
required for methane and isobutane with a final delivery pressure of
15 psi (103 kPa). Clean tubing must be used. It must be solvent-
washed and nitrogen-dried. The bulkhead connector at the rear of the
instrument is a 1/8 in. Swagelok fitting.
The use of commercial gas purifiers for reagent gas is recommended.
Safety Requirements
Gas Delivery
Lines
Copper tubing that is free of grease, oil and organic material must
always be used with the Clarus MS on all gas lines, except
ammonia reagent gas. Ammonia reagent gas requires stainless
steel.
Solvent-washed copper tubing must be used to avoid
contamination of the Gas Chromatograph. Suitable solvents are
acetone or dichloromethane (do not use if negative chemical
ionization is planned) followed by methanol. Clean helium or
nitrogen should be used to blow any residual solvent from the
tubing. Cap all unused tubing.
Care must be taken not to kink or overstress the gas delivery
lines. Strain relief consisting of two one inch coils of tubing
should be installed at every gas line connection.
Gas Cylinders
All gas cylinders should be firmly clamped to a suitable surface.
Gas cylinders should be located outside of the laboratory
whenever possible, and should always be stored and operated in a
vertical position.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......